Can You Run an Oxygen Concentrator on a UPS?

Yes, and Philips' own EverFlo manual recommends a UPS for anyone who needs the concentrator to keep running through mains interruptions. But a UPS buys minutes, not hours: at the 280 to 350 W most 5 L/min units are rated, a common 1500 VA tower lasts about 13 to 17 minutes. Its job is to ride through flickers and give time to switch to backup oxygen.

On this page
  1. What the manuals say concentrators draw
  2. The two rules that look like they conflict
  3. Runtime by UPS size
  4. Alarms: what you will and will not hear
  5. Transfer time and waveform
  6. Setting it up
  7. Where this answer stops applying
  8. Frequently asked questions

A stationary oxygen concentrator runs a compressor continuously, and the 5 L/min machines most people have at home are rated at 280 to 350 W. A UPS will carry one, but the figure to plan around is minutes, and the useful question is what those minutes are for.

This page covers the power engineering, not medical advice. Backup oxygen and how long someone can go without the concentrator are decisions for the prescriber and the oxygen supplier. Our medical equipment guide covers the wider outage plan.

What the manuals say concentrators draw

Makers state these figures differently: some give an average, some a figure at one flow setting, one a maximum.

Published power ratings for home oxygen concentrators (120 V, 60 Hz unless noted)
ConcentratorFlow rangePublished ratingSource
Inogen At Home (GS-100)1 to 5 L/min100 W at 2 L/min; 275 W at 5 L/minInogen
Invacare Perfecto20.5 to 5 L/min280 W at 3 L/min; 3 A rated inputUser manual
CAIRE Companion 50.5 to 5 L/min285 W at 2 L/min; 350 W maximumSpec sheet
AirSep VisionAire 51 to 5 L/min290 W at all flow ratesSpec sheet
Drive DeVilbiss 525DS0.5 to 5 L/min310 W average; 275 W at 1.2 L/min and belowDrive product page
Philips Respironics EverFlo0.5 to 5 L/min350 WUser manual
Invacare Platinum 10L2 to 10 L/min585 W typical; 4.9 A rated inputUser manual

The watt figure sets runtime. The rated input current sets how big a UPS has to be: the Perfecto2's 3 A at 120 V is 360 VA, and the Platinum 10L's 4.9 A is about 590 VA. Both UPS ratings must clear those figures with room to spare; see VA vs watts.

The Inogen At Home spans 100 to 275 W across its settings, so its runtime depends heavily on the prescription. For any machine, a plug-in meter at the prescribed flow gives the real number; see how to measure power draw.

The two rules that look like they conflict

One common answer says a concentrator must go straight into the wall, because the manual forbids extension cords. Another says a UPS is the obvious backup. The EverFlo manual contains both.

Its warnings say: "Do not use extension cords or electrical adapters." Its electromagnetic immunity table, in the row for voltage dips and interruptions, says: "If the user of the device requires continued operation during power mains interruptions, it is recommended that the device be powered from an uninterruptible power supply or a battery." Same manual, same machine.

The Invacare manuals show what the cord rule is about. Both the Perfecto2 and the Platinum 10L list "Insufficient power at outlet" as a cause of the System Failure alarm, and the fix they give is "DO NOT use extension cords. Move to another electrical outlet or circuit." The concern is voltage at the plug. Neither Invacare manual mentions a UPS for or against.

Bottom line

The no-extension-cord rule and a UPS are compatible when the concentrator plugs directly into a battery outlet on a UPS sized for it, and the UPS plugs directly into a wall receptacle. What the manuals rule out is anything in between: cords, adapters, and the power strip people often put under a UPS (see plugging a UPS into a power strip). If your manual says something different from the ones cited here, follow your manual and ask the supplier.

Runtime by UPS size

These estimates use the model behind our UPS runtime calculator: new sealed lead-acid batteries, a Peukert exponent of 1.25 referenced to the 20-hour rating, 88% inverter efficiency and 8 W of inverter overhead. Battery sizes are typical for each class.

Estimated runtime by concentrator rating and UPS class (new batteries, room temperature)
UPS class (battery)100 W (At Home, setting 2)280 W (Perfecto2)310 W (525DS)350 W (EverFlo)585 W (Platinum 10L)
600 VA / 360 W (1 x 12 V 7 Ah)18 min5 min5 mintoo close to limitover rating
850 VA / 510 W (1 x 12 V 9 Ah)24 min7 min6 min5 minover rating
1000 VA / 600 W (2 x 12 V 7 Ah)42 min12 min11 min9 mintoo close to limit
1500 VA / 900 W (2 x 12 V 9 Ah)58 min17 min15 min13 min7 min
1500 VA / 1350 W (4 x 12 V 7 Ah)1 h 40 min29 min26 min22 min12 min
2200 VA (2 x 12 V 18 Ah)2 h 17 min40 min36 min31 min16 min
3000 VA (4 x 12 V 18 Ah)5 h 25 min1 h 36 min1 h 24 min1 h 13 min39 min

Here is the arithmetic for an EverFlo on a 1500 VA tower. The battery supplies the concentrator plus overhead, divided by inverter efficiency:

Battery draw = (350 W + 8 W) / 0.88 = 407 W, or 17.0 A at 24 V
Runtime = 20 h x (9 Ah / (17.0 A x 20 h))^1.25 = about 0.21 h, or 13 minutes

Dividing 216 Wh by 350 W says 37 minutes. Lead-acid batteries deliver far less than their label at this rate, nearly 2C for a 9 Ah block, so the real figure is about a third of the naive one; why UPS runtime is not linear explains the effect. It also explains why the 4-battery 1500 VA unit beats its extra capacity: 56% more watt-hours, but about 70% more runtime, because each battery carries half the current.

The rows marked "too close to limit" are loads at 97% or more of the UPS's watt rating. A compressor that the UPS can barely hold on a cool morning may trip it on a hot one, so treat 80% of the watt rating as the working ceiling; see UPS load percentage.

Our analysis: what the minutes are for

Thirteen minutes sounds useless, but without a UPS every flicker stops the compressor. The manuals describe a warm-up after switching on: the EverFlo "typically takes 10 minutes to reach oxygen purity specifications," and Invacare gives about 30 minutes for the Perfecto2 and Platinum 10L to reach maximum purity, though both say the patient can breathe from the device during that time. A UPS turns a blink into nothing at all. For a longer outage, the same minutes are the calm window to move to cylinders before the concentrator's alarm, rather than after it. Planning a UPS to carry the night is the mistake: eight hours at 280 W needs about 2,550 Wh from the battery, which is power station or generator territory.

Alarms: what you will and will not hear

On a UPS, the concentrator never sees the outage, so its own power failure alarm stays silent. The UPS's on-battery beep is the only warning that the clock is running, which is a strong reason not to silence it; UPS beeping explains the patterns. When the UPS battery is exhausted, the concentrator's alarm takes over:

  • Invacare Perfecto2 and Platinum 10L: both list a "Battery Free Power Loss Alarm," and the troubleshooting table describes "Short Beeps, long pause" with no lights. The manuals say the concentrator "will briefly alarm signaling the patient to switch to their backup source of oxygen."
  • Philips EverFlo: a continuous audible alarm with no LEDs lit: "Often this indicates that the device is not plugged in or there is a power failure." The manual says to connect to a backup oxygen source.
  • Drive DeVilbiss 525DS: Drive lists audible and visual alarms including "power fail."

"Briefly" matters: if nobody hears the UPS beep, the concentrator's own alarm may be the only sound. Keep the backup oxygen where the user can reach it without help.

Transfer time and waveform

Line-interactive units commonly switch to battery in 2 to 6 ms (see UPS transfer time). The EverFlo's immunity table includes a dip of more than 95% for half a cycle, 8.3 ms at 60 Hz, but it states test levels, not a promise of uninterrupted running, so verify the transfer with your own machine. An online double-conversion UPS has no transfer at all; UPS types explained covers the trade.

No manual cited here sets an output waveform, but motors run hotter and noisier on stepped output. Choose pure sine if you are buying for this job.

Setting it up

  1. Read the rating, both watts and rated current, and confirm with a meter at the prescribed flow if you can.
  2. Choose the size from the table, keeping the concentrator under about 80% of the UPS watt rating. For a 280 to 350 W unit that means 850 VA / 510 W or larger; for a Platinum 10L, 1500 VA / 900 W or larger.
  3. Plug the concentrator directly into a battery outlet, not a surge-only outlet, and the UPS directly into a wall receptacle. No cords, adapters or strips.
  4. Keep the UPS to the concentrator. Anything else on it comes straight out of the minutes.
  5. Test it. With the concentrator running at the prescribed flow and the backup oxygen ready, switch off the UPS's input and time it. Repeat yearly, because lead-acid capacity fades; see when to replace a UPS battery.

Where this answer stops applying

  • Flows above 5 L/min. Invacare warns that patients on more than 5 L/min "may be at an increased risk for serious injury or death in the event of failure" and tells them to discuss it with the supplier before use.
  • 230 V regions. The figures move a little: the EverFlo's 230 V models are rated 320 W at 60 Hz and under 300 W at 50 Hz, and Drive's 230 V Compact 525 about 290 W at 2 L/min and 312 W at 5 L/min.
  • Outages longer than the battery. HHS's ASPR TRACIE notes that home oxygen users "often go to health care facilities to charge oxygen concentrators during power outages or obtain refills of personal oxygen cylinders during disasters." Plan that with the supplier, not with a bigger UPS.
  • Shared loads. A CPAP on the same UPS adds its draw; rerun the numbers.

Frequently asked questions

How many watts does a home oxygen concentrator use?

Most 5 L/min stationary units are rated between about 280 and 350 W: Invacare Perfecto2 280 W at 3 L/min, CAIRE Companion 5 285 W at 2 L/min, AirSep VisionAire 5 290 W, Drive DeVilbiss 525DS 310 W average, Philips EverFlo 350 W. The Inogen At Home is the low outlier at 100 W on setting 2. A 10 L/min Invacare Platinum 10L is rated 585 W typical.

How long will a 1500 VA UPS run an oxygen concentrator?

With two 12 V 9 Ah batteries, the usual fit, about 13 minutes at 350 W, 17 minutes at 280 W and 7 minutes at 585 W, on new batteries at room temperature. An Inogen At Home on setting 2, at 100 W, gets close to an hour. Older batteries give less; test your own setup by unplugging the UPS while the concentrator runs.

My manual says no extension cords. Does that rule out a UPS?

Not in itself. The EverFlo manual carries both rules: no extension cords or electrical adapters, and a recommendation to use an uninterruptible power supply or battery if you need operation through power interruptions. Invacare lists an extension cord among the causes of "insufficient power at outlet". The point is full voltage at the plug. Plug the concentrator directly into a correctly sized UPS, with no cord or strip between.

Will my concentrator alarm when the power goes out if it is on a UPS?

No. The concentrator keeps running on UPS power, so its own power failure alarm stays silent until the UPS battery is exhausted. The UPS's on-battery beep is your only early warning, so leave it audible. When the UPS shuts down, the concentrator's power loss alarm sounds (Invacare's is battery free), which is the signal to be on backup oxygen already.

Does an oxygen concentrator need a pure sine wave UPS?

None of the manuals cited here states a waveform requirement. A concentrator is built around a compressor motor, and motors run hotter and noisier on the stepped output of simulated sine units. If you are buying a UPS for a concentrator, a pure sine line-interactive model removes the question at modest extra cost.

How much battery would it take to run a concentrator overnight?

Far more than a UPS holds. Eight hours at 280 W is 2,240 Wh at the outlet, about 2,550 Wh from a battery after 88% inverter efficiency, more than ten times the 216 Wh in a typical 1500 VA tower. That is the scale of a large portable power station or a generator, chosen with your oxygen supplier.

Sources and further reading

  1. Philips Respironics: EverFlo / EverFlo Q User Manual (electrical specifications, warnings, EMC immunity table)
  2. Invacare: Perfecto2 Oxygen Concentrator User Manual, IRC5PO2, IRC5PO2W, IRC5P, IRC5PW (1143482-H-00)
  3. Invacare: Platinum 10L Oxygen Concentrator User Manual (1118353-M-00)
  4. Drive DeVilbiss: 5 Liter Oxygen Concentrator 525DS specifications
  5. Drive DeVilbiss: Compact 525 5 Litre Oxygen Concentrator spec sheet (230 V)
  6. CAIRE: Companion 5 stationary oxygen concentrator spec sheet (ML-CONC0142)
  7. CAIRE: AirSep VisionAire 5 spec sheet (ML-CONC0082)
  8. Inogen: Inogen At Home energy savings calculator (100 W at 2 L/min, 275 W at 5 L/min)
  9. HHS ASPR TRACIE: Utility Failure Tip Sheet: Oxygen